Ozone Monitor
Introduction
BARC has developed an ozone monitor for measuring ozone concentrations at ppb levels (10-9 v/v ratio) in the ambient atmosphere. It is based on chemiluminescent reaction of ozone with ethylene supplied from a cylinder. Light produced in this reaction in the region of 300-500 nm is measured by a sensitive photomultiplier tube and the signal is processed and displaced on the meter and recorded by a recorder or data processor. Light produced in the reaction is specific to ozone and other pollutants do not interfere.
Specifications
Measuring ranges | : | 0-100 ppb |
: | 0-1 ppm | |
: | 0- 10 ppm | |
Minimum detection limit | : | 5 ppb |
Response time (10% to 90% rise) | : | 5 sec |
Reproducibility> | : | +1% |
Linearity | : | +1% |
Output | : | 0 - 1 volt (for air quality data processor) |
Power requirements> | : | 100 watt at 230 v, 50 Hz |
Applications
Ozone monitor can be used for measuring ambient levels of ozone in polluted urban atmospheres and near high voltage power distribution centres.
-
For further information please contact
Head, Technology Transfer & Collaboration Division,
Bhabha Atomic Research Centre,
Trombay, Mumbai- 400 085
Tel : 091-022-25593897
Fax : 091-022-25505151 / 25519613
Email : headttcd@barc.gov.in