Cardiac Output Monitor

Introduction

 

Cardiac Output Monitor reveals pump performance of the heart. The board provides Cardiac Output signal in analog form which can be displayed on standard Oscilloscope. It will form front end module to Patient Monitoring systems useful for the monitoring of stroke volume and cardiac output in patients. It is based on the principle of electrical impedance plethysmography and is non-invasive in nature. 

This method has been reported to be accurate in majority of subjects except those with intracardiac shunts or valvular regurgitation and at present is the only non-invasive method for continuous cardiac output monitoring.

 

Specifications 

 

Principle

:

Electrical impedance measurement by constant current method.

Mode of operation

:

Calibration/Patient

Configuration

:

Tetrapolar

Sensor

:

Spot electrodes

Carrier current

:

Sinusoidal at 48 KHz frequency and 4 mA amplitude

Measurement range

:

Impedance 5 to 75 W 

Rate of change of impedance

:

-1.5 to +3.5 W/sec

Stroke volume

:

10 ml to 150 ml

Cardiac output

:

0.5 litres to 20 litres

Patient isolation

:

3000 Volts

Power input

:

230 VAC +/- 10%, 50 Hz, 100 W

Output

:

On Oscilloscope 

 

Applications

 

This can be used for OEM application for continuous monitoring of stroke volume and cardiac output in critically ill patients in Intensive Care Units and Intensive Cardiac Care Units, with hardware addition.

 